Load Cell Calibration Calculator
A load cell gives a small voltage that grows with the load. This calculator works out that voltage from the rated output in mV/V, finds how many converter counts one unit of load gives, and fits a calibration line through readings of known weights.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use this calculator
- Choose what you want: the output for a load (or the load for an output), the resolution with a converter, or a calibration from known loads.
- Enter the rated output (for example 2 mV/V), the excitation voltage and the rated capacity from the label or certificate of the cell. Type the capacity and the loads in one unit.
- For a calibration, put known weights on the cell, note each reading, and type the pairs into the rows. Add a reading to turn it into a load.
- Choose Calculate.
The formulas
- Output = rated output (mV/V) × excitation (V) × load ÷ capacity. At full capacity the output is rated output × excitation.
- Load from an output = output ÷ (rated output × excitation) × capacity.
- Counts per unit of load = (full-scale output ÷ capacity) × gain ÷ LSB, with LSB = converter range ÷ 2^N, the rule of the Adc Resolution Calculator. The resolution is one count turned into load.
- Calibration line: reading = zero reading + factor × load, fitted by least squares. With two points it goes through both. Then load = (reading − zero reading) ÷ factor.
Worked examples
A 50 kg cell with 2 mV/V at 5 V excitation: the full-scale output is 2 × 5 = 10 mV, so a 20 kg load gives 10 × 20 ÷ 50 = 4 mV.
The same cell with a gain of 128 and a 24-bit converter that sees a 5 V range: one step is 5 ÷ 16,777,216 = 0.298 µV, the full-scale output after the gain is 1.28 V, so the cell gives about 85,899 counts per kg, or 0.0000116 kg (0.0116 g) per count before noise.
Calibration with 0, 10, 20 and 50 kg giving readings of 8,430, 93,150, 177,800 and 432,100: the line has a factor of 8,473.5 per kg and a zero reading of 8,400. A reading of 260,000 is (260,000 − 8,400) ÷ 8,473.5 = 29.693 kg. The point furthest from the line is off by 70 counts, which is 0.0083 kg.
Limits you should know
- The cell and the electronics are taken as linear. The rated output, zero offset, nonlinearity, creep and temperature effects of a real cell are in its data sheet and calibration certificate.
- Noise sets the real resolution. A 24-bit converter rarely gives more than about 16 to 20 bits that are free of noise.
- Do not load a cell beyond its rated capacity.
- This tool does not decide whether a scale is legal for trade. That needs a scale approved under the rules of your country.
- For the strain and bridge formulas behind a load cell, see the Strain Gauge Calculator.