Conductivity to TDS Calculator

Estimate total dissolved solids (TDS) from a conductivity reading, or the conductivity from a TDS value. You choose the conversion factor, because no single factor fits every water.

Enter your values

Estimate total dissolved solids (TDS, in mg/L or ppm) from a conductivity reading with a factor you choose.

Your values

The reading of your meter.

Most TDS and EC meters correct the reading to 25 °C themselves (ATC). A reading that is not corrected changes by about 2% per °C.

TDS in mg/L = k × EC in µS/cm. About 0.55 to 0.75 for natural waters; meters use 0.5, 0.64 or 0.7. There is no single right value.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose Conductivity to TDS or TDS to conductivity.
  2. Enter the reading and choose its unit: µS/cm, mS/cm (EC in hydroponics), dS/m, S/m or CF.
  3. Choose the factor k, or pick one from the menu. If your meter does not correct for temperature, choose At the water temperature and type the temperature and the coefficient.
  4. Choose Calculate. The result also shows the TDS for other common factors, so you can see how much the choice matters.

The formulas

  • TDS (mg/L) = k × EC at 25 °C (µS/cm), and so EC = TDS ÷ k. ppm here means mg/L, which is right for fresh water.
  • The factor k depends on the salts in the water. It is about 0.55 to 0.75 for most natural waters. TDS meters use a fixed factor: 0.5 (the "500 scale", based on sodium chloride), 0.64 (the "640 scale") or 0.7 (the "700 scale", used by many hydroponic meters).
  • Temperature: conductivity rises by about 2% for each °C. A linear correction gives EC25 = ECt ÷ (1 + α (t − 25 °C)), where α is the coefficient in % per °C that you type.
  • Your own factor: k = TDS measured by drying and weighing ÷ EC at 25 °C of the same water.
  • Units: 1 mS/cm = 1,000 µS/cm = 1 dS/m = 0.1 S/m = 10 CF. Resistivity is 1 ÷ conductivity: 1 µS/cm = 1 MΩ·cm.

Worked examples

Tap water. A reading of 500 µS/cm at 25 °C with k = 0.65 gives TDS = 0.65 × 500 = 325 mg/L (ppm). With k = 0.5 it would be 250 mg/L, and with k = 0.7 it would be 350 mg/L.

Hydroponics. EC 1.4 mS/cm is 1,400 µS/cm: 700 ppm on a meter with factor 0.5, and 980 ppm on a meter with factor 0.7.

Not corrected. A meter without correction reads 450 µS/cm in water at 20 °C. With α = 2% per °C: EC25 = 450 ÷ (1 + 0.02 × (20 − 25)) = 450 ÷ 0.9 = 500 µS/cm.

Pure water. 18.2 MΩ·cm is 1 ÷ 18.2 = 0.0549451 µS/cm.

Limits you should know

  • This is an estimate. Only a dried and weighed sample measures TDS itself. Dissolved matter without a charge (sugar, silica, many organic substances) does not show in conductivity.
  • The factors in the menu are examples. Use the factor of your meter, or find your own with Find the factor for your water.
  • The linear temperature correction is close for natural waters between about 0 and 30 °C. Very pure water and strong brines follow other curves.
  • For the salinity of seawater from conductivity (the Practical Salinity Scale 1978), use the Salinity Calculator.