Toolvito

Works on this device

Laboratory Centrifuge Rcf Calculator

Plan a centrifuge run with the data of your rotor. Find the g-force (RCF) at the top, the middle and the bottom of the tube, the speed for an RCF, the k-factor and pelleting time, and the time or speed that gives the same result in another rotor.

Enter your values

The g-force across the tube in a rotor, from the speed, or the speed for an RCF a protocol asks for.

Your values

Revolutions per minute of the rotor.

From the centre of the rotor to the top of the liquid in the tube, from the rotor data sheet.

From the centre of the rotor to the bottom of the tube, from the rotor data sheet.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose what you want to work out, for example RCF at the minimum, average and maximum radius.
  2. Type the minimum radius (r min, the top of the liquid) and the maximum radius (r max, the bottom of the tube) from the data sheet of your rotor. The values offered by Try an example are examples only.
  3. Type the speed in rpm, or choose to start from the RCF a protocol asks for.
  4. Choose Calculate.

The formulas

  • RCF = 1.11824 × 10⁻⁵ × radius in cm × rpm². The number is (2π ÷ 60)² × 0.01 ÷ 9.80665, with standard gravity 9.80665 m/s². Lab sheets often round it to 1.118 × 10⁻⁵.
  • Average radius r avg = (r min + r max) ÷ 2.
  • k-factor (clearing factor) = 2.53303 × 10¹¹ × ln(r max ÷ r min) ÷ rpm². The number is 10¹³ ÷ (3600 × (2π ÷ 60)²). Manuals often round it to 2.53 × 10¹¹.
  • k-factor at another speed = printed k × (speed of the printed k ÷ your speed)².
  • Time to pellet in hours = k ÷ S, for a particle with a sedimentation coefficient of S Svedberg units.
  • Time in another rotor: t₁ ÷ k₁ = t₂ ÷ k₂, so t₂ = t₁ × k₂ ÷ k₁.
  • Same RCF in another rotor: speed₂ = speed₁ × √(radius₁ ÷ radius₂).

Worked example

A rotor with r min = 3.7 cm and r max = 8.4 cm at 13,000 rpm has r avg = 6.05 cm and gives 6,992 × g at r min, 11,433 × g at r avg and 15,875 × g at r max. Its k-factor at this speed is 2.53303 × 10¹¹ × ln(8.4 ÷ 3.7) ÷ 13,000² = 1,229. A particle of 1,000 S pellets in 1,229 ÷ 1,000 = 1.23 hours, about 74 minutes.

For 10,000 × g at r max the same rotor needs √(10,000 ÷ (1.11824 × 10⁻⁵ × 8.4)) = 10,318 rpm.

A protocol runs 30 minutes in a rotor with k = 1,229. A rotor with k = 439 does the same pelleting in 30 × 439 ÷ 1,229 = 10.7 minutes.

A protocol says 4,000 rpm in a rotor of 15 cm radius, which is 2,684 × g. A rotor of 9.5 cm radius needs 4,000 × √(15 ÷ 9.5) = 5,026 rpm for the same g-force.

Limits you should know

  • Use the radii and k-factors of your own rotor, from its manual. This page keeps no rotor data.
  • The k-factor rule holds for pelleting from r min to r max. It does not hold for gradients, for runs that stop early, or when the tubes are not full.
  • A pelleting time from k ÷ S is for a liquid like water at the temperature the S value is for. A thicker or colder liquid takes longer.
  • Never run a rotor or tubes above their rated speed, and balance the tubes. Follow the manual of your centrifuge.
  • For a quick conversion between rpm and RCF at one radius, use the Rpm to Rcf Calculator.