Toolvito

Works on this device

Viable Cell Percentage Calculator

Find the percentage of live (viable) cells from a dye-exclusion count such as trypan blue. If you counted live and dead cells in each square of a hemocytometer, the page also gives the live and the total cells per mL.

Enter your values

The percentage of live cells from a trypan blue (or similar) count.

Your values

Cells that keep the dye out: clear, bright cells with trypan blue.

Cells that take up the dye: blue cells with trypan blue.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose Viability from live and dead counts for totals, or Live and dead counts per square for a hemocytometer.
  2. Type the live cells (clear, not stained) and the dead cells (blue with trypan blue), or the live cells and all cells.
  3. For counts per square, choose the kind of square and type the dilution factor: 2 when you mixed equal volumes of cells and trypan blue.
  4. Choose Calculate.

The formulas

  • Viability = live cells ÷ (live cells + dead cells) × 100.
  • Dead cells in % = 100 − viability.
  • Live cells per mL = average live cells per square × dilution factor × 10,000 (for the large squares of an improved Neubauer chamber, 0.1 µL each) or × 250,000 (for the small centre squares, 0.004 µL each).
  • All cells per mL = live cells per mL + dead cells per mL.

Worked example

You counted 180 clear cells and 20 blue cells. Viability = 180 ÷ (180 + 20) × 100 = 90%.

In four large squares you counted 45, 48, 43 and 46 live cells and 5, 4, 6 and 3 dead cells, after mixing 1:1 with trypan blue. That is 182 live and 18 dead cells, a viability of 182 ÷ 200 × 100 = 91%. Live cells per mL = 182 ÷ 4 × 2 × 10,000 = 910,000, and all cells per mL = 200 ÷ 4 × 2 × 10,000 = 1,000,000.

Limits you should know

  • Dye exclusion shows a broken cell membrane. A cell that is dying but still has a whole membrane counts as live, so the test can read high.
  • Count within a few minutes of adding the dye. Live cells also take it up with time, and the viability then reads low.
  • Clumps and debris make the count less exact. A count of a few cells has a large random error, which the page shows for the live count.
  • For plain cell counts, use the Hemocytometer Counting Calculator. To dilute or seed the live cells, use the Cell Dilution Calculator and the Cell Concentration Calculator.

Related tools