Measurement Bias Calculator

Bias is how far the average reading of an instrument is from the true value. Measure a reference several times, paste the readings, and see the bias, whether it could be zero, and how it changes across the range of the instrument.

Enter your values

Measure a reference part or standard several times. The page gives the bias, its confidence interval and a t-test.

Your values

The accepted value of the thing measured, from a calibration or a better instrument, in the unit of the readings.

If you give it, the bias is also shown as a percentage of this width. Use the full width of the tolerance, or six standard deviations of the process, as your procedure says.

Readings of the reference

A column or a row of numbers copied from a spreadsheet. Use a decimal point, and leave out headings and units.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose Bias from repeated readings of a reference for one reference part or standard.
  2. Type the reference value. Then type the readings, or paste a column from a spreadsheet and choose Read into rows.
  3. Choose the level of confidence, and choose Calculate.
  4. For a linearity study, choose Linearity: bias across the range and give one row for each reading: the reference value, then the reading.

The formulas

  • Bias = mean of the readings − reference value. A positive bias means the instrument reads high.
  • Sample standard deviation s = √(sum of (reading − mean)² ÷ (n − 1)). It divides by n − 1.
  • t statistic = bias ÷ (s ÷ √n), with n − 1 degrees of freedom. The p-value is two-sided.
  • Confidence interval of the bias = bias ± t value × s ÷ √n. If zero is inside the interval, the readings do not show a bias at that level.
  • Linearity: the bias of every reading is fitted with the line bias = intercept + slope × reference value by least squares. The slope and the intercept are each tested against zero with n − 2 degrees of freedom.
  • Linearity as a percentage = size of the slope × 100. With a process variation, linearity = size of the slope × process variation.

Worked example

A reference of 10 mm is measured ten times: 10.02, 10.01, 10.03, 10.02, 10.00, 10.02, 10.01, 10.03, 10.02 and 10.01 mm. The mean is 10.017 mm, so the bias is 10.017 − 10 = 0.017 mm.

The sample standard deviation is 0.0095 mm, so s ÷ √n = 0.0095 ÷ √10 = 0.003 mm, and t = 0.017 ÷ 0.003 = 5.667 with 9 degrees of freedom. The two-sided p-value is 0.0003.

The Student t value for 95% and 9 degrees of freedom is 2.262. The 95% confidence interval of the bias is 0.017 ± 2.262 × 0.003, from 0.0102 to 0.0238 mm. Zero is outside it, so the bias is statistically significant.

Limits you should know

  • The reference value must be known much better than the instrument can measure. Its uncertainty is not included in the result.
  • The tests assume independent readings that are roughly normal. In a linearity study the scatter is taken to be the same over the range.
  • A bias that is not significant is not proof of no bias. With few readings or much scatter, only a large bias can be found.
  • A significant bias can still be small enough for the job, and one that is not significant can be too large. Compare the bias and its interval with your tolerance.
  • The page gives no accept or reject limit. Your quality system or your customer sets it.
  • Repeatability and reproducibility studies are in the Measurement System Analysis Calculator.
  • Results guide decisions about instruments. An accredited laboratory and the written procedure decide.