Works on this device
Metrology Calculator
Measurement arithmetic for the inspection room and the workshop. Set a sine bar, build a gauge block stack, check a vernier or micrometer reading, or work out the uncertainty of a measurement. Choose a calculation and enter what you know.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose the button below them.
Your result
How it was worked out
How to use this calculator
- Choose what you want to calculate, for example Sine bar: gauge block height and angle.
- Enter your values. Each box has a short note that says what it needs.
- For repeated readings, type them in the rows or paste a column from a spreadsheet and choose Read into rows.
- Choose Calculate. You can copy the result, print it, or download the tables as a CSV file.
What each calculation does
- Sine bar. Height of gauge blocks = length of the bar × sin(angle). The other way: angle = arcsin(height ÷ length). The result also shows how much the angle changes for 1 µm of height.
- Gauge block stack. Finds the fewest blocks of a set that add up to a size, with each block used once. You can change the rows to match your own set.
- Vernier caliper reading. Reading = main scale mark before the vernier zero + vernier line × least count − zero error. The least count is one main scale division ÷ the number of vernier divisions.
- Micrometer reading. Reading = sleeve + thimble line × value of one line + vernier line × one tenth of a line − zero error.
- Type A uncertainty. Mean, sample standard deviation s (it divides by n − 1), standard uncertainty of the mean s ÷ √n with n − 1 degrees of freedom, and the expanded uncertainty with a coverage factor from Student’s t.
- Uncertainty budget. Each value ÷ its divisor × its sensitivity coefficient is a contribution. The combined standard uncertainty is the root of the sum of the squared contributions. The effective degrees of freedom come from the Welch-Satterthwaite formula.
- Thermal correction. Length at 20 °C = measured length × (1 + α of the scale × ΔT of the scale) ÷ (1 + α of the part × ΔT of the part).
- Test uncertainty ratio. Width of the tolerance ÷ (2 × expanded uncertainty), and acceptance limits with a guard band.
Worked examples
Sine bar. A 5 inch sine bar set to 30°: height = 5 × sin(30°) = 5 × 0.5 = 2.5 in.
Gauge blocks. To build 1.3745 in from an 81 block inch set: 0.1005 + 0.124 + 0.15 + 1 = 1.3745 in. That is 4 blocks, and no group of 3 blocks from the set makes this size.
Vernier. A caliper that reads to 0.02 mm shows 23 mm before the vernier zero, and vernier line 21 meets a main scale line: 23 + 21 × 0.02 = 23.42 mm.
Type A. Ten readings with a mean of 10.017 mm and a sample standard deviation of 0.0095 mm: standard uncertainty of the mean = 0.0095 ÷ √10 = 0.003 mm. For 95% and 9 degrees of freedom, k = 2.26, so the expanded uncertainty is 0.0068 mm.
Budget. Contributions of 0.0032, 0.002, 0.00029 and 0.00115 mm combine to √(0.0032² + 0.002² + 0.00029² + 0.00115²) = 0.00396 mm.
Limits you should know
- The vernier and micrometer calculations work with numbers. They are not a caliper drawn on the screen that you can slide.
- The gauge block sets in the menu are common sets. Sets differ between makers, so check the rows against the list in your own box.
- The uncertainty budget is first-order propagation for inputs that do not depend on each other (GUM, JCGM 100:2008). Correlated inputs and strongly curved formulas need more than this table.
- The coverage factor comes from Student’s t at the effective degrees of freedom, rounded down to a whole number. Above 1,000,000 degrees of freedom the normal distribution is used.
- Expansion coefficients are values you enter. The example values in the notes are typical handbook figures for trying the tool, not data for your material.
- The 4 to 1 ratio and the guard band rule are requirements of standards and customers, not laws of measurement. Use what your own quality system sets.
- Results guide engineering and quality decisions. They do not replace a calibrated instrument, a written procedure or an accredited laboratory.