Measurement arithmetic for the inspection room and the workshop. Set a sine bar, build a gauge block stack, check a vernier or micrometer reading, or work out the uncertainty of a measurement. Choose a calculation and enter what you know.

Enter your values

The height of gauge blocks that sets a sine bar or a sine plate to an angle, or the angle that a height gives.

Your values

The distance between the centres of the two rolls. It is marked on the bar, for example 5 in, 10 in, 100 mm or 200 mm.

You can type decimal degrees here, such as 30.5, and leave minutes and seconds at 0.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose the button below them.

How to use this calculator

  1. Choose what you want to calculate, for example Sine bar: gauge block height and angle.
  2. Enter your values. Each box has a short note that says what it needs.
  3. For repeated readings, type them in the rows or paste a column from a spreadsheet and choose Read into rows.
  4. Choose Calculate. You can copy the result, print it, or download the tables as a CSV file.

What each calculation does

  • Sine bar. Height of gauge blocks = length of the bar × sin(angle). The other way: angle = arcsin(height ÷ length). The result also shows how much the angle changes for 1 µm of height.
  • Gauge block stack. Finds the fewest blocks of a set that add up to a size, with each block used once. You can change the rows to match your own set.
  • Vernier caliper reading. Reading = main scale mark before the vernier zero + vernier line × least count − zero error. The least count is one main scale division ÷ the number of vernier divisions.
  • Micrometer reading. Reading = sleeve + thimble line × value of one line + vernier line × one tenth of a line − zero error.
  • Type A uncertainty. Mean, sample standard deviation s (it divides by n − 1), standard uncertainty of the mean s ÷ √n with n − 1 degrees of freedom, and the expanded uncertainty with a coverage factor from Student’s t.
  • Uncertainty budget. Each value ÷ its divisor × its sensitivity coefficient is a contribution. The combined standard uncertainty is the root of the sum of the squared contributions. The effective degrees of freedom come from the Welch-Satterthwaite formula.
  • Thermal correction. Length at 20 °C = measured length × (1 + α of the scale × ΔT of the scale) ÷ (1 + α of the part × ΔT of the part).
  • Test uncertainty ratio. Width of the tolerance ÷ (2 × expanded uncertainty), and acceptance limits with a guard band.

Worked examples

Sine bar. A 5 inch sine bar set to 30°: height = 5 × sin(30°) = 5 × 0.5 = 2.5 in.

Gauge blocks. To build 1.3745 in from an 81 block inch set: 0.1005 + 0.124 + 0.15 + 1 = 1.3745 in. That is 4 blocks, and no group of 3 blocks from the set makes this size.

Vernier. A caliper that reads to 0.02 mm shows 23 mm before the vernier zero, and vernier line 21 meets a main scale line: 23 + 21 × 0.02 = 23.42 mm.

Type A. Ten readings with a mean of 10.017 mm and a sample standard deviation of 0.0095 mm: standard uncertainty of the mean = 0.0095 ÷ √10 = 0.003 mm. For 95% and 9 degrees of freedom, k = 2.26, so the expanded uncertainty is 0.0068 mm.

Budget. Contributions of 0.0032, 0.002, 0.00029 and 0.00115 mm combine to √(0.0032² + 0.002² + 0.00029² + 0.00115²) = 0.00396 mm.

Limits you should know

  • The vernier and micrometer calculations work with numbers. They are not a caliper drawn on the screen that you can slide.
  • The gauge block sets in the menu are common sets. Sets differ between makers, so check the rows against the list in your own box.
  • The uncertainty budget is first-order propagation for inputs that do not depend on each other (GUM, JCGM 100:2008). Correlated inputs and strongly curved formulas need more than this table.
  • The coverage factor comes from Student’s t at the effective degrees of freedom, rounded down to a whole number. Above 1,000,000 degrees of freedom the normal distribution is used.
  • Expansion coefficients are values you enter. The example values in the notes are typical handbook figures for trying the tool, not data for your material.
  • The 4 to 1 ratio and the guard band rule are requirements of standards and customers, not laws of measurement. Use what your own quality system sets.
  • Results guide engineering and quality decisions. They do not replace a calibrated instrument, a written procedure or an accredited laboratory.