Detection Limit Calculator
Find the limit of detection (LOD) of an analytical method by the method you choose: from a calibration line as in ICH Q2, from replicate blank readings, or from the signal-to-noise ratio. The limit of quantification (LOQ) by the same method is shown beside it.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use this calculator
- Choose the method your laboratory uses: a calibration line, a slope and a standard deviation you already have, blank readings, or signal-to-noise.
- Type your values. For a line, type each standard in a row (you can paste them); for blanks, type the readings separated by commas.
- Under More options you can change the factors (3.3, 3, 10) if your method names others, and type the highest LOD your method allows to check against it.
- Choose Calculate. The result names the method and the factors, and shows the LOD, the LOQ and the working.
The methods
- Calibration line (ICH Q2): LOD = 3.3 × σ ÷ S and LOQ = 10 × σ ÷ S. S is the slope of the line. σ is the residual standard deviation of the line (s_y/x) or the standard deviation of its y-intercept, s_a = s_y/x × √(1/n + x̄² ÷ Σ(x − x̄)²). Choose which σ and say so in your report.
- Slope and standard deviation you have (ICH Q2): the same formulas, with σ for example the standard deviation of the responses of blank samples.
- Blank readings: the response at the LOD is the blank mean + 3 standard deviations of the blanks, and at the LOQ the mean + 10 standard deviations. With the slope of your line these become the concentrations 3 × SD ÷ S and 10 × SD ÷ S.
- Signal-to-noise: the ratio of a low sample is S/N = 2H ÷ h (European Pharmacopoeia 2.2.46, USP <621>; H the peak height, h the peak-to-peak noise of the baseline) or H ÷ noise as your method measures it. The LOD is the concentration with S/N = 3 and the LOQ the one with S/N = 10.
Worked examples
Calibration line. Standards of 0, 0.05, 0.1, 0.2, 0.4 and 0.8 µg/mL give responses of 1.1, 3.0, 6.2, 10.1, 21.3 and 40.4. The line has a slope S = 49.5945 and a residual standard deviation s_y/x = 0.5383. So LOD = 3.3 × 0.5383 ÷ 49.5945 = 0.0358 µg/mL and LOQ = 10 × 0.5383 ÷ 49.5945 = 0.1085 µg/mL. With the standard deviation of the intercept (0.3018) instead, LOD = 0.0201 µg/mL.
Blank readings. Ten blanks have a mean of 0.0126 and a standard deviation of 0.002221. The response at the LOD is 0.0126 + 3 × 0.002221 = 0.01926. With a slope of 0.025 per µg/mL, LOD = 3 × 0.002221 ÷ 0.025 = 0.2665 µg/mL and LOQ = 0.8884 µg/mL.
Signal-to-noise. A 0.5 µg/mL standard gives a peak 1,250 high on a baseline with a peak-to-peak noise of 95. S/N = 2 × 1,250 ÷ 95 = 26.32, so LOD = 0.5 × 3 ÷ 26.32 = 0.057 µg/mL and LOQ = 0.19 µg/mL.
Limits you should know
- Different methods give different limits for the same data, as the examples show. Always report the method, the σ and the factors with the limit.
- A limit worked out from data must be confirmed by measuring samples at or near it, as ICH Q2 asks. The tool cannot do that part.
- Use standards near the expected limit. A line through high standards often has a larger σ and gives a limit that is too high.
- The highest limit your method allows comes from your own method, customer or regulator. The tool holds no official limits; the example value is only for trying the tool.
- The US EPA method detection limit (MDL) procedure, with spiked samples and Student's t, is not built in.
- For the limit of quantification first, use the Quantification Limit Calculator. It has the same methods and gives the same numbers.