Type the amount of standard added to each portion of your sample and the signal you measured. The tool fits the best straight line, extends it to zero signal and gives the concentration of the sample with its standard deviation.

Enter your values

You know the concentration each addition gives in the measured solution. The line through the portions is extended to zero signal.

Your values

The unit of the added concentrations or of the standard. The answer is in the same unit.

Spiked portions

One row for each portion of the sample: the concentration of standard added to it (as it is in the measured solution), then its signal. The first row is usually the sample with nothing added.

One portion on each line: the amount added, then the signal, such as 5, 0.41. Leave out headings and units.

More options

If the sample was diluted in each portion (10 mL made up to 50 mL is 5), the result is multiplied by this. Leave 1 if the added concentrations are for the sample itself.

If you type it, it is taken away from every signal. The method needs signals that are zero when no analyte is there.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose how you made the additions: as concentrations you know, as volumes of a standard made up to one final volume, or as volumes added to the sample itself.
  2. Type one row for each portion: the amount added, then the signal. The first row is usually the sample with nothing added. You can also paste the rows, one portion on each line.
  3. Type the unit of concentration and, for volumes, the concentration of the standard and the volumes. More options has a dilution factor, a blank signal and the confidence level.
  4. Choose Calculate. The result shows the concentration of the sample, its standard deviation and interval, the line and a chart of it.

How the method works

In standard addition, known amounts of the analyte are added to portions of the sample itself, so the standards and the sample share the same matrix. The signal rises in a straight line with the amount added. The tool fits that line, signal = a + b × x, by least squares.

The line meets zero signal at x = −a ÷ b. The distance from there to zero is the concentration that was already in the measured solution: c = a ÷ b.

The standard deviation of c comes from the scatter of the points about the line (Miller and Miller): s(c) = (s_y/x ÷ b) × √(1/n + ȳ² ÷ (b² × Σ(x − x̄)²)), where s_y/x is the residual standard deviation, n the number of portions, ȳ the mean signal and x̄ the mean amount added. The confidence interval is c ± t × s(c), with Student's t at n − 2 degrees of freedom.

  • Made up to one volume: each flask holds V(sample) of sample and V(std) of a standard of concentration C(std), filled to V(final). The amount added is C(std) × V(std) ÷ V(final), and the sample holds c × V(final) ÷ V(sample).
  • Added to the sample itself: each addition makes the volume larger and dilutes the analyte. Each signal is multiplied by (V0 + V) ÷ V0 to undo this dilution, and the amount added is C(std) × V ÷ V0. The x-intercept then gives the concentration of the sample directly (Harris, Quantitative Chemical Analysis).

Worked example

Portions of a sample get 0, 5, 10, 15, 20, 25 and 30 ng/mL of silver and read 0.32, 0.41, 0.52, 0.60, 0.70, 0.77 and 0.89. The best line is signal = 0.321786 + 0.0186429 × x, with r² = 0.99755.

The concentration is 0.321786 ÷ 0.0186429 = 17.26 ng/mL. The residual standard deviation is 0.01092, so s(c) = 0.748 ng/mL, and with t = 2.5706 for 5 degrees of freedom the 95% interval is 17.26 ± 1.92 ng/mL.

With volumes: 10 mL of sample with 0, 0.5, 1, 1.5 and 2 mL of a 100 ng/mL standard, each made up to 25 mL, read 0.241, 0.338, 0.442, 0.539 and 0.641. The flasks hold 4.795 ng/mL from the sample, so the sample holds 4.795 × 25 ÷ 10 = 11.99 ng/mL.

Limits you should know

  • The answer is found by extending the line past the measured points, so it is less precise than a reading inside a calibration line. Use several portions, and additions that reach about two to three times the signal of the sample.
  • The signal must be zero when no analyte is there. Subtract the blank (or type it under More options). Standard addition corrects for a matrix that changes the slope; it does not correct for a background signal that adds to every reading.
  • The signal must rise in a straight line over the whole range. Look at the differences of the points from the line in the table.
  • The tool does not convert concentration units: the answer is in the unit you type.
  • For an ordinary calibration line with separate standards, use the Absorbance to Concentration Calculator. For the recovery of a single spike, use the Sample Recovery Calculator.