Transmittance to Absorbance Calculator
Convert transmittance (%T or T) to absorbance, or absorbance to transmittance. Type one value or a whole list. You can also start from the two light readings, I and I0.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Convert.
Your result
How it was worked out
How to use this calculator
- Choose Transmittance to absorbance, Absorbance to transmittance, or light readings if you have the detector values I and I0.
- For transmittance, choose whether your values are percents (%T, such as 45.2) or fractions (T, such as 0.452).
- Type one value or a list separated by commas. Choose Convert. The table shows T, %T and A for every value, and you can download it as a CSV file.
The formulas
Transmittance is the share of the light that passes through the sample: T = I ÷ I0, where I is the light through the sample and I0 the light through the blank. As a percent, %T = 100 × T.
Absorbance is A = −log10 T. With a percent, the same formula is A = 2 − log10(%T). The other way round, T = 10^(−A) and %T = 100 × 10^(−A).
Absorbance has no unit. Each step of 1 in absorbance means ten times less light: A = 1 is 10% T, A = 2 is 1% T and A = 3 is 0.1% T.
| %T | T | A |
|---|---|---|
| 100% | 1 | 0 |
| 50% | 0.5 | 0.301 |
| 10% | 0.1 | 1 |
| 1% | 0.01 | 2 |
| 0.1% | 0.001 | 3 |
Worked examples
Percent to absorbance. A reading of 45.2% T gives A = 2 − log10(45.2) = 2 − 1.655138 = 0.344862. A reading of 85% T gives A = 0.070581.
Absorbance to percent. An absorbance of 0.345 gives T = 10^(−0.345) = 0.451856, so 45.1856% of the light passes. An absorbance of 0.05 gives 89.1251% T.
Light readings. With I = 452 and I0 = 1,000, T = 0.452 and A = log10(1,000 ÷ 452) = 0.344862.
Limits you should know
- Absorbance and transmittance are linked by the formula alone; the tool needs no other value. To go from a concentration to an absorbance, use the Beer Lambert Calculator, which shows T beside A. To read concentrations from standards, use the Absorbance to Concentration Calculator.
- 100 − %T is the light that does not reach the detector. It is not the absorbance, and it also holds light that is reflected or scattered.
- Above an absorbance of about 2, less than 1% of the light passes, and stray light makes the reading unreliable. Dilute the sample or use a shorter cuvette.
- A transmittance above 100% gives a negative absorbance. Near zero this is reading noise; larger values point to a wrong blank. The tool takes absorbances from −1 to 10.