Toolvito

Works on this device

Absorbance to Concentration Calculator

Type the concentrations and absorbances of your standards and the absorbances of your samples. The tool fits the best straight line and reads the concentration of each sample from it.

Enter your values

Measure standards of known concentration, fit the best straight line, and read the samples from it.

Your values

The unit your standards are in. The answers are in the same unit.

One or more absorbances, separated by commas, spaces or new lines.

Standards

One row for each standard: its known concentration, then the absorbance you measured. A blank (concentration 0) may be one of them.

One standard on each line: the concentration, then the absorbance, such as 2, 0.152. Leave out headings and units.

More options

If you diluted a sample before you measured it (1 mL made up to 10 mL is 10), the result is multiplied by this.

Only when your absorbances are blank-corrected and the response is known to start at zero.

If you type it, it is taken away from every absorbance, standards and samples alike. Leave it empty if your readings are already blank-corrected.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Measure a set of standards of known concentration, and a blank if you have one.
  2. Type each standard in a row: its concentration, then its absorbance. You can also paste them, one standard on each line. Type the unit of the concentrations, such as mg/L or µM.
  3. Type the absorbances of your samples, separated by commas. If you diluted the samples, open More options and type the dilution factor.
  4. Choose Calculate. The result shows the line, r², the concentration of each sample, and a warning for any sample outside the range of the standards.

How it works

By the Beer-Lambert law, absorbance rises in a straight line with concentration. The tool fits the line A = a + b × c through your standards by least squares: it chooses the slope b and the intercept a that make the sum of the squared differences of the absorbances as small as possible.

A sample with absorbance A then has the concentration c = (A − a) ÷ b, times any dilution factor. The standard deviation of that concentration comes from the scatter of the standards about the line (ISO 8466-1): s(c) = (s ÷ b) × √(1 + 1/n + (A − mean A)² ÷ (b² × Σ(c − mean c)²)).

r², the coefficient of determination, shows how close the standards lie to a straight line. A value near 1 is good, but it does not prove the line is right for your samples. Look at the differences of the standards from the line, and keep your samples inside the range of the standards.

Worked example

Standards of 0, 2, 4, 6, 8 and 10 mg/L read 0.003, 0.152, 0.298, 0.451, 0.597 and 0.749. The best line is A = 0.002286 + 0.074543 × c, with r² = 0.99997.

A sample that reads 0.380 then holds (0.380 − 0.002286) ÷ 0.074543 = 5.067 mg/L, with a standard deviation of about 0.026 mg/L. A sample that reads 0.520 holds 6.945 mg/L.

With only one standard of 5 mg/L that reads 0.372, a sample reading 0.451 holds about 5 × 0.451 ÷ 0.372 = 6.062 mg/L.

Limits you should know

  • Keep samples inside the range of your standards. Outside it, the line is extended and the answer may be wrong; dilute the sample and measure again.
  • Use the same cuvette type, wavelength and blank for the standards and the samples.
  • Above an absorbance of about 1.5 to 2, stray light bends the line. Dilute such samples.
  • A line forced through zero, and a single standard, both assume that zero concentration gives zero absorbance. Use them only when that is true for your method.
  • If you know the molar absorptivity ε instead, use the Beer Lambert Calculator. To turn %T readings into absorbance, use the Transmittance to Absorbance Calculator.

Related tools