Calculates on this device. File exports run online.
Lab Replicate Statistics Calculator
Type or paste replicate results of one sample to get the mean, standard deviation, relative standard deviation, the confidence interval and Grubbs’ test for an outlier. You can also work out repeatability and intermediate precision from results of several runs, or the bias and recovery against a reference value.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use the replicate statistics calculator
- Choose Statistics of replicates and type the results (or paste a column of numbers).
- For precision, choose Repeatability and intermediate precision and give each result with its run (day, analyst or series).
- For trueness, give the results and the reference value of the material, with its uncertainty if you have it.
- The box under the result works the statistics out again on the server with normality tests and plots.
The formulas
- Standard deviation s = √(Σ(x − x̄)² ÷ (n − 1)); RSD = s ÷ |x̄|; standard error s ÷ √n; interval x̄ ± t s ÷ √n.
- Grubbs: G = max |x − x̄| ÷ s against G_crit = ((n − 1) ÷ √n) √(t² ÷ (n − 2 + t²)), t at α ÷ (2n) with n − 2 degrees of freedom; ISO 5725-2 calls a result a straggler above the 5% value and an outlier above the 1% value.
- Precision by one-way ANOVA: s_r² = MS within runs, s_L² = (MS between − MS within) ÷ n̄, s_R² = s_r² + s_L²; limits r = 2.8 s_r and R = 2.8 s_R (ISO 5725-6).
Worked example
The six results 10.12, 10.08, 10.15, 10.11, 10.09 and 10.31 have a mean of 10.143, s = 0.0852 and RSD 0.84%. Grubbs’ G for 10.31 is 1.955, between the 5% value 1.887 and the 1% value 1.973: a straggler, to be checked but not simply dropped.
Limits you should know
- The tests assume normally distributed results. Remove a result only when you find a cause, not on a test alone.
- Intermediate precision here is within one laboratory; reproducibility needs an interlaboratory study.