Calibration Curve Calculator

Type the concentrations and responses of your standards, then the mean response of each unknown sample. You get the calibration equation y = a + b x with its standard errors and r², the limits of detection and quantitation, and the concentration of each unknown with its standard error and confidence interval.

Enter your values

Your values

Standards

Each standard: its known concentration and the measured response (absorbance, peak area …). Repeat a concentration for replicate standards.

One standard per line: the concentration, then the response, separated by a tab, comma, semicolon or space.

Unknown samples

The mean response of each unknown and how many readings it is the mean of.

More options

Only when a blank truly gives zero response; the intercept is otherwise kept.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

Line fitted on the server

Line fitted on the server

Uses the standards and unknowns above. The server fits the line again with statsmodels (with parameter intervals and residuals), reads the unknowns back and gives their uncertainty from the parameter covariance, and draws the line with the residuals.

The calculator above works on your device. When you run a task in this section, Toolvito receives the values and result you submit to make the file or run the check. Download the result within one hour after the job finishes. Choose Delete files now to request earlier removal. The hourly cleanup normally removes files within one to two hours after completion.

How to use the calibration curve calculator

  1. Type each standard: its concentration and the measured response (absorbance, peak area, signal). Or paste two columns.
  2. Add each unknown sample with its mean response and the number of readings it is the mean of.
  3. Press Calculate. Check the residuals and the plot before you trust the line.
  4. The box under the result fits the same line on the server and gives the uncertainty another way.

The formulas (Miller and Miller 2010)

  • Slope b = Σ(x − x̄)(y − ȳ) ÷ Σ(x − x̄)², intercept a = ȳ − b x̄, s_y/x = √(Σ(y − ŷ)² ÷ (n − 2)).
  • Unknown: x0 = (y0 − a) ÷ b with s_x0 = (s_y/x ÷ b) √(1/m + 1/n + (y0 − ȳ)² ÷ (b² Σ(x − x̄)²)) and the interval x0 ± t s_x0.
  • LOD = 3.3 s_y/x ÷ b and LOQ = 10 s_y/x ÷ b (ICH Q2(R1)).

Worked example

Standards at 0, 2, 4, 6, 8 and 10 with absorbances 0.009, 0.158, 0.301, 0.472, 0.577 and 0.739 give y = 0.0133 + 0.07254 x with r = 0.9988. An unknown with a mean absorbance of 0.420 from 3 readings is 5.607 ± 0.143 (standard error), with a 95% interval of 5.21 to 6.00.

Limits you should know

  • A straight, unweighted line: when the spread of the responses grows with the concentration, a weighted fit is better. Stay inside the range of the standards.