Calculates on this device. File exports run online.
Calibration Curve Calculator
Type the concentrations and responses of your standards, then the mean response of each unknown sample. You get the calibration equation y = a + b x with its standard errors and r², the limits of detection and quantitation, and the concentration of each unknown with its standard error and confidence interval.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use the calibration curve calculator
- Type each standard: its concentration and the measured response (absorbance, peak area, signal). Or paste two columns.
- Add each unknown sample with its mean response and the number of readings it is the mean of.
- Press Calculate. Check the residuals and the plot before you trust the line.
- The box under the result fits the same line on the server and gives the uncertainty another way.
The formulas (Miller and Miller 2010)
- Slope b = Σ(x − x̄)(y − ȳ) ÷ Σ(x − x̄)², intercept a = ȳ − b x̄, s_y/x = √(Σ(y − ŷ)² ÷ (n − 2)).
- Unknown: x0 = (y0 − a) ÷ b with s_x0 = (s_y/x ÷ b) √(1/m + 1/n + (y0 − ȳ)² ÷ (b² Σ(x − x̄)²)) and the interval x0 ± t s_x0.
- LOD = 3.3 s_y/x ÷ b and LOQ = 10 s_y/x ÷ b (ICH Q2(R1)).
Worked example
Standards at 0, 2, 4, 6, 8 and 10 with absorbances 0.009, 0.158, 0.301, 0.472, 0.577 and 0.739 give y = 0.0133 + 0.07254 x with r = 0.9988. An unknown with a mean absorbance of 0.420 from 3 readings is 5.607 ± 0.143 (standard error), with a 95% interval of 5.21 to 6.00.
Limits you should know
- A straight, unweighted line: when the spread of the responses grows with the concentration, a weighted fit is better. Stay inside the range of the standards.