Calibration Uncertainty Calculator
Work out the uncertainty of one calibration point. Type the certificate value of your reference standard, the readings and the resolution of the device under test, and any further components. The page gives the error of indication, the expanded uncertainty and a statement for the certificate.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use this calculator
- Choose One calibration point from repeated readings. Type the value of the reference standard, its expanded uncertainty and the coverage factor from its certificate.
- Type the readings of the device under test, or paste a column from a spreadsheet and choose Read into rows.
- Type the resolution of the device: one display step, or the smallest step that you read.
- Add a row for each further component, such as temperature or the drift of the standard. The menu Lists of further components fills in common names for one kind of calibration. The values stay empty for you to fill.
- Choose Calculate. You can print the result or download the budget as a CSV file.
The method
- Error of indication = mean of the readings of the device − value of the reference standard.
- Reference standard: standard uncertainty = expanded uncertainty from its certificate ÷ its coverage factor k.
- Resolution of the device under test: standard uncertainty = one step ÷ √12. This is a rectangular distribution over plus and minus half a step.
- Repeatability: s ÷ √n for the mean of n readings, where s is the sample standard deviation (it divides by n − 1). It has n − 1 degrees of freedom.
- Further components: value ÷ divisor of its distribution: √3 rectangular, √6 triangular, √2 U-shaped, k for a certificate value.
- Combined standard uncertainty u_c = √(sum of the squared contributions). Expanded uncertainty U = k × u_c.
- The coverage factor is the Student t value for the level of confidence at the effective degrees of freedom (Welch-Satterthwaite formula), or a fixed 2 or 3.
- The method is that of the Guide to the Expression of Uncertainty in Measurement (GUM). For calibration laboratories it is described in the document EA-4/02.
Worked example
A reference standard of 10 mm has an expanded uncertainty of 0.004 mm with k = 2, so its standard uncertainty is 0.002 mm. The device under test shows steps of 0.01 mm: 0.01 ÷ √12 = 0.0029 mm.
Ten readings have a mean of 10.017 mm and s = 0.0095 mm. Repeatability of the mean = 0.0095 ÷ √10 = 0.003 mm, with 9 degrees of freedom. Temperature is given as a limit of ± 0.002 mm: 0.002 ÷ √3 = 0.00115 mm.
u_c = √(0.002² + 0.0029² + 0.003² + 0.00115²) = 0.0048 mm. The effective degrees of freedom are 57.09, so k for 95.45% is 2.04 and U = 0.0097 mm.
The error of indication is 10.017 − 10 = 0.017 mm. The result is 0.0170 ± 0.0097 mm.
Where the neighbouring jobs are
- A measurement that is not a calibration, with the same kind of budget: Measurement Uncertainty Calculator.
- The uncertainty that comes from a display step or a scale division, with other reading rules: Instrument Resolution Calculator.
- The test uncertainty ratio against a tolerance and guard bands: Metrology Calculator.
- When the next calibration is due: Calibration Interval Planner.
Limits you should know
- The budget is for one calibration point. Work out each point of a range by itself.
- The components are taken as independent, and the propagation is first-order.
- Resolution and repeatability are both counted. Some procedures count only the larger of the two. Type 0 for the resolution to leave it out.
- The lists of further components are names that often belong to a calibration of that kind. They are not complete and they are not taken from your procedure.
- The statement is a draft. Your accreditation body and your procedure decide the wording of a certificate.
- This is arithmetic, not professional advice. It does not replace a written procedure or an accredited laboratory.