Toolvito

Works on this device

Microscope Calibration Calculator

Calibrate the scale in your eyepiece (a reticle or graticule) against a stage micrometer, one objective at a time, and turn eyepiece divisions into micrometres. You can also work out the micrometres per pixel of a microscope camera.

Enter your values

Find the length of one division of the eyepiece scale with each objective, from a stage micrometer.

Your values

Printed on the stage micrometer. A scale of 1 mm in 100 parts has divisions of 0.01 mm, which is 10 µm.

Type it to see how long the whole eyepiece scale is at the specimen with each objective.

Calibration for each objective

For each objective, line up the two scales and count how many stage divisions match how many eyepiece divisions, over as long a stretch as you can.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to calibrate an eyepiece scale

  1. Put the stage micrometer on the stage and focus on it with the first objective.
  2. Turn the eyepiece so that its scale lies along the stage scale. Line up a mark of each scale at the left.
  3. Find the place furthest to the right where a mark of the eyepiece scale lines up with a mark of the stage scale again. Count the divisions of each scale between the two places.
  4. Type the objective and the two counts in a row of the table. Do the same for each objective.
  5. Type the length of one stage division (often 10 µm) and choose Calculate.

The formulas

  • One eyepiece division = number of stage divisions × length of one stage division ÷ number of eyepiece divisions.
  • Size of an object = eyepiece divisions across it × length of one eyepiece division for that objective.
  • Micrometres per pixel from a known length = known length ÷ its length in pixels.
  • Micrometres per pixel from the camera = pixel size of the camera × binning ÷ (objective × camera adapter × any other factor).

Worked example

With the 4× objective, 50 stage divisions of 10 µm line up with 20 eyepiece divisions. One eyepiece division is 50 × 10 ÷ 20 = 25 µm. With the 40× objective, 10 stage divisions line up with 41 eyepiece divisions: 10 × 10 ÷ 41 = 2.44 µm.

A cell 4 eyepiece divisions long at 40× is 4 × 2.44 = 9.76 µm long.

A photo of 100 µm of the stage micrometer is 412.5 pixels long, so one pixel is 100 ÷ 412.5 = 0.2424 µm, and there are 4.125 pixels per micrometre. A camera with 3.45 µm pixels on a 40× objective and a 1× adapter gives 3.45 ÷ 40 = 0.08625 µm per pixel.

Limits you should know

  • Calibrate each objective on its own. A new eyepiece, a magnification changer or another tube factor needs a new calibration.
  • A calibration from a stage micrometer is more exact than one from the numbers engraved on the parts, which are nominal.
  • A camera calibration holds for one objective, adapter, binning and image size. An image that is made smaller later has larger pixels.
  • To put a scale bar on an image with your micrometres per pixel, use the Microscopy Scale Bar Generator. For magnification, use the Microscope Magnification Calculator.