Microscope Calibration Calculator
Calibrate the scale in your eyepiece (a reticle or graticule) against a stage micrometer, one objective at a time, and turn eyepiece divisions into micrometres. You can also work out the micrometres per pixel of a microscope camera.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to calibrate an eyepiece scale
- Put the stage micrometer on the stage and focus on it with the first objective.
- Turn the eyepiece so that its scale lies along the stage scale. Line up a mark of each scale at the left.
- Find the place furthest to the right where a mark of the eyepiece scale lines up with a mark of the stage scale again. Count the divisions of each scale between the two places.
- Type the objective and the two counts in a row of the table. Do the same for each objective.
- Type the length of one stage division (often 10 µm) and choose Calculate.
The formulas
- One eyepiece division = number of stage divisions × length of one stage division ÷ number of eyepiece divisions.
- Size of an object = eyepiece divisions across it × length of one eyepiece division for that objective.
- Micrometres per pixel from a known length = known length ÷ its length in pixels.
- Micrometres per pixel from the camera = pixel size of the camera × binning ÷ (objective × camera adapter × any other factor).
Worked example
With the 4× objective, 50 stage divisions of 10 µm line up with 20 eyepiece divisions. One eyepiece division is 50 × 10 ÷ 20 = 25 µm. With the 40× objective, 10 stage divisions line up with 41 eyepiece divisions: 10 × 10 ÷ 41 = 2.44 µm.
A cell 4 eyepiece divisions long at 40× is 4 × 2.44 = 9.76 µm long.
A photo of 100 µm of the stage micrometer is 412.5 pixels long, so one pixel is 100 ÷ 412.5 = 0.2424 µm, and there are 4.125 pixels per micrometre. A camera with 3.45 µm pixels on a 40× objective and a 1× adapter gives 3.45 ÷ 40 = 0.08625 µm per pixel.
Limits you should know
- Calibrate each objective on its own. A new eyepiece, a magnification changer or another tube factor needs a new calibration.
- A calibration from a stage micrometer is more exact than one from the numbers engraved on the parts, which are nominal.
- A camera calibration holds for one objective, adapter, binning and image size. An image that is made smaller later has larger pixels.
- To put a scale bar on an image with your micrometres per pixel, use the Microscopy Scale Bar Generator. For magnification, use the Microscope Magnification Calculator.