Works on this device
Microscope Resolution Calculator
Find the smallest distance between two points that a microscope objective can show apart. Type the numerical aperture (NA) of the objective and the wavelength of the light. The page gives the Abbe limit, the Rayleigh limit and the Rayleigh limit with the condenser, and the NA you need for a detail.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use this calculator
- Choose Resolution of an objective, or Numerical aperture needed for a detail.
- Type the numerical aperture of the objective. It is engraved after the magnification, such as 40×/0.65.
- Type the wavelength of the light, or choose an example wavelength from the list. For fluorescence, use the emission wavelength of the dye.
- Optional: type the NA of the condenser for the Rayleigh limit with the condenser.
- Choose Calculate. A smaller distance means finer detail.
The formulas
- Abbe limit: d = λ ÷ (2 × NA). Ernst Abbe showed that an objective can only pass lines that are at least this far apart.
- Rayleigh limit: d = 0.61 × λ ÷ NA. Two points are just seen apart when the middle of one blur spot lies on the first dark ring of the other.
- Rayleigh limit with the condenser: d = 1.22 × λ ÷ (NA of the objective + NA of the condenser). Closing the condenser iris lowers its NA and makes d larger.
- Numerical aperture: NA = n × sin θ, where n is the refractive index of the medium between the specimen and the objective (air 1.00, water 1.33, immersion oil about 1.52) and θ is half the angle of the cone of light the objective takes in.
- NA needed for a detail d: NA = λ ÷ (2 d) by Abbe, or NA = 0.61 λ ÷ d by Rayleigh.
Worked example
A 40× objective with an NA of 0.65 in green light of 550 nm: the Abbe limit is 550 ÷ (2 × 0.65) = 423 nm, and the Rayleigh limit is 0.61 × 550 ÷ 0.65 = 516 nm. With a condenser at NA 0.9 the Rayleigh limit is 1.22 × 550 ÷ (0.65 + 0.9) = 433 nm.
To see details 0.25 µm apart in 550 nm light, you need an NA of 550 ÷ (2 × 250) = 1.1 by Abbe, or 0.61 × 550 ÷ 250 = 1.342 by Rayleigh. That needs an oil immersion objective.
Limits you should know
- These limits hold for ideal optics, correct focus and a thin specimen with good contrast. A real image is often less sharp.
- The example wavelengths are examples. White light has many wavelengths; 550 nm is a common choice for the middle of it.
- Magnification does not add resolution. The Microscope Magnification Calculator shows the useful magnification of an objective, 500 to 1,000 × NA as a rule of thumb.
- For the width of the field you see, use the Microscope Field of View Calculator.
- Super-resolution and electron microscopes work past these limits by other means; this page does not cover them.