Works on this device
Microscope Magnification Calculator
Find the total magnification of a compound light microscope: objective × eyepiece. You can also find a missing objective or eyepiece, work out the real size of a cell from a drawing or photo, and check the useful magnification of an objective.
Enter your values
Check these values
Your values are worked out in this browser. They are not sent to a server.
Your result will show here
Fill in the values, then choose Calculate.
Your result
How it was worked out
How to use this calculator
- Choose what you want to work out, for example Total magnification.
- Type the magnification engraved on the objective (such as 40×) and on the eyepiece (such as 10×).
- If the light path has another lens, type its factor, such as 1.5 for a magnification changer. Otherwise leave 1.
- Optional: type the numerical aperture (NA) of the objective, such as 0.65 from 40×/0.65, to check the useful magnification.
- Choose Calculate. You can copy the result or download it as a CSV file.
The formulas
- Total magnification = objective magnification × eyepiece magnification × any other factor. A 40× objective with a 10× eyepiece gives 400×.
- A missing lens: objective = total ÷ (eyepiece × factor), and eyepiece = total ÷ (objective × factor).
- Magnification of an image = size in the image ÷ real size, with both sizes in the same unit. So real size = image size ÷ magnification. This works for a drawing, a photo, a screen and a scale bar.
- Useful magnification, a rule of thumb: from about 500 × NA to 1,000 × NA. Above it the image only gets larger ("empty magnification"); the detail is set by the resolution of the objective.
Worked example
A 40× objective and a 10× eyepiece give 40 × 10 = 400×. The objective is marked 40×/0.65, so the useful range is 500 × 0.65 = 325× to 1,000 × 0.65 = 650×, and 400× is inside it.
A cell is 24 mm long in a photo taken at 400×. Its real size is 24 mm ÷ 400 = 0.06 mm, which is 60 µm.
With the objectives 4×, 10×, 40× and 100× and a 10× eyepiece, the total magnifications are 40×, 100×, 400× and 1,000×. Low power is 40× and high power (oil) is 1,000×.
Limits you should know
- Magnification only makes the image larger. How much detail you can see depends on the numerical aperture and the light: see the Microscope Resolution Calculator.
- For the width of the circle you see, use the Microscope Field of View Calculator. To measure with an eyepiece scale or to find the size of a camera pixel, use the Microscope Calibration Calculator.
- A camera on a microscope does not see through the eyepiece, so its image scale comes from the objective and the camera adapter, not from this total. The calibration calculator works it out.
- The useful range comes from a rule of thumb, not from a law. Your eyes, the specimen and the light also matter.
- A printed photo or a screen adds its own enlargement. Measure the image you have, or use a scale bar, rather than the number on the microscope.