Find the total magnification of a compound light microscope: objective × eyepiece. You can also find a missing objective or eyepiece, work out the real size of a cell from a drawing or photo, and check the useful magnification of an objective.

Enter your values

The magnification of a compound light microscope from its objective and eyepiece.

Your values

Engraved on the objective, such as 4×, 10×, 40× or 100×.

Engraved on the eyepiece (the ocular), such as 10× or 15×.

1 for none. Type the factor of a magnification changer (such as 1.5), a tube factor printed on the head (such as 1.25) or an auxiliary lens of a stereo microscope (such as 0.5).

Engraved after the magnification, such as 40×/0.65. Type it to check the useful magnification.

Your values are worked out in this browser. They are not sent to a server.

Your result will show here

Fill in the values, then choose Calculate.

How to use this calculator

  1. Choose what you want to work out, for example Total magnification.
  2. Type the magnification engraved on the objective (such as 40×) and on the eyepiece (such as 10×).
  3. If the light path has another lens, type its factor, such as 1.5 for a magnification changer. Otherwise leave 1.
  4. Optional: type the numerical aperture (NA) of the objective, such as 0.65 from 40×/0.65, to check the useful magnification.
  5. Choose Calculate. You can copy the result or download it as a CSV file.

The formulas

  • Total magnification = objective magnification × eyepiece magnification × any other factor. A 40× objective with a 10× eyepiece gives 400×.
  • A missing lens: objective = total ÷ (eyepiece × factor), and eyepiece = total ÷ (objective × factor).
  • Magnification of an image = size in the image ÷ real size, with both sizes in the same unit. So real size = image size ÷ magnification. This works for a drawing, a photo, a screen and a scale bar.
  • Useful magnification, a rule of thumb: from about 500 × NA to 1,000 × NA. Above it the image only gets larger ("empty magnification"); the detail is set by the resolution of the objective.

Worked example

A 40× objective and a 10× eyepiece give 40 × 10 = 400×. The objective is marked 40×/0.65, so the useful range is 500 × 0.65 = 325× to 1,000 × 0.65 = 650×, and 400× is inside it.

A cell is 24 mm long in a photo taken at 400×. Its real size is 24 mm ÷ 400 = 0.06 mm, which is 60 µm.

With the objectives 4×, 10×, 40× and 100× and a 10× eyepiece, the total magnifications are 40×, 100×, 400× and 1,000×. Low power is 40× and high power (oil) is 1,000×.

Limits you should know

  • Magnification only makes the image larger. How much detail you can see depends on the numerical aperture and the light: see the Microscope Resolution Calculator.
  • For the width of the circle you see, use the Microscope Field of View Calculator. To measure with an eyepiece scale or to find the size of a camera pixel, use the Microscope Calibration Calculator.
  • A camera on a microscope does not see through the eyepiece, so its image scale comes from the objective and the camera adapter, not from this total. The calibration calculator works it out.
  • The useful range comes from a rule of thumb, not from a law. Your eyes, the specimen and the light also matter.
  • A printed photo or a screen adds its own enlargement. Measure the image you have, or use a scale bar, rather than the number on the microscope.